Glide force is a clinical property that comes from the barrel and plunger fit, which means it is fixed before the first part is moulded.

Glide force is what a clinician feels, and it comes from the clearance between barrel and plunger together with the surface finish of both. Neither of those can be adjusted by changing process parameters after the tool exists.

That makes glide force a tooling specification rather than a process one. It has to be stated as a target at the design stage, along with the acceptable range, and verified on moulded samples rather than calculated from drawings.

The same logic applies to luer taper fit. Two tapers have to agree within hundredths of a millimetre for a connection to seal without locking, and both are set in steel long before any process engineer sees the part.

Specifications

Set byBarrel and plunger clearance and surface finish
Fixed atTool design, not process adjustment
Verified onMoulded samples, not drawings